ProductGuru Recalls

Official recall record

Mixing Bowl And Spatula, 20- Pack

Health Canada — Medical Devices · 2026-04-14 · CanadaType II
Issuing authority
Health Canada — Medical Devices
Published
2026-04-14
Country of alert
Canada
Alert level
Type II
Alert type
recall

What the authority reported

Zimmer Biomet is conducting a lot specific medical device recall removal for the mixing bowl and spatula. Ten complaints have been received identifying issues at the time of use related to the package seal, including incomplete seals, wrinkles in the seals or peeling seals. Seal integrity testing conducted internally showed potential sterility breach in 7.5% of the tested units that had visual seal nonconformities. The devices are distributed in 20-pack boxes. The defectiveness was discovered on November 19, 2025. Issue was identified through the complaints handling process.

What to do

Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.

Verify at the source: Health Canada — Medical Devices — official record →
Record last checked against the registry on 2026-05-31.

More recalls: Health Canada — Medical Devices · Canada · all filters