ProductGuru Recalls

Official recall record

Monnal™ T60 Advanced

Health Canada — Medical Devices · 2026-04-30 · CanadaType II
Issuing authority
Health Canada — Medical Devices
Published
2026-04-30
Country of alert
Canada
Alert level
Type II
Alert type
recall

What the authority reported

Air Liquide Medical Systems has identified a manufacturing non-conformity that affects a limited and specific number of Monnal T60 and Monnal T60 advanced devices. It concerns the failure to fit washers beneath the screw head that secures the inspiratory unit of the device. The absence of a washer calls into question the conformity of the Monnal T60 units concerned with the requirements of standard EM 13718-1 (Requirements for medical devices used in air ambulances) and standard EN 1789+A1 (Requirements for medical devices used in road ambulances).

What to do

Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.

Verify at the source: Health Canada — Medical Devices — official record →
Record last checked against the registry on 2026-05-31.

More recalls: Health Canada — Medical Devices · Canada · all filters

ProductGuru Recalls republishes official safety information with attribution. We are not the issuing authority and this page is not advice. If this product is a medical device or medication, speak to a doctor or pharmacist. Found an error? Report it — corrections are published.