ProductGuru Recalls

Official recall record

directCHECK™ Whole Blood Control

Health Canada — Medical Devices · 2026-05-14 · CanadaType II
Issuing authority
Health Canada — Medical Devices
Published
2026-05-14
Country of alert
Canada
Alert level
Type II
Alert type
recall

What the authority reported

Instances of consistently elevated results were identified when using directCHECK™ Whole Blood Control DCGLR‑2, lot A6DLA001. A process error led to testing the lot on the wrong instrument platform, which resulted in the creation and release of incorrect Quality Control (QC) ranges on the performance range card distributed with this lot.

What to do

Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.

Verify at the source: Health Canada — Medical Devices — official record →
Record last checked against the registry on 2026-05-31.

More recalls: Health Canada — Medical Devices · Canada · all filters

ProductGuru Recalls republishes official safety information with attribution. We are not the issuing authority and this page is not advice. If this product is a medical device or medication, speak to a doctor or pharmacist. Found an error? Report it — corrections are published.