Exoplan

Health Canada — Medical Devices · 2026-05-20 · CanadaType II
Issuing authority
Health Canada — Medical Devices
Published
2026-05-20
Country of alert
Canada
Alert level
Type II
Alert type
recall

What the authority reported

During internal testing of libraries an error was identified within specific Straumann surgical protocol libraries used with Exoplan version 3.1. Under certain implant/sleeve/workflow selections, the protocol output displays an incorrect instrument combination in the final drilling steps. In two consecutive protocol steps an incorrect drill key is displayed, which may result in a drilling depth up to 2 mm deeper than intended. This may lead to unintended injury depending on anatomical conditions and clinical use. The deviation is limited to specific implant/sleeve/workflow selections and does not affect all protocol configurations.

What to do

Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.

Verify at the source: Health Canada — Medical Devices — official record →
Record last checked against the registry on 2026-05-31.