Official recall record
Exoplan
- Issuing authority
- Health Canada — Medical Devices
- Published
- 2026-05-20
- Country of alert
- Canada
- Alert level
- Type II
- Alert type
- recall
What the authority reported
During internal testing of libraries an error was identified within specific Straumann surgical protocol libraries used with Exoplan version 3.1. Under certain implant/sleeve/workflow selections, the protocol output displays an incorrect instrument combination in the final drilling steps. In two consecutive protocol steps an incorrect drill key is displayed, which may result in a drilling depth up to 2 mm deeper than intended. This may lead to unintended injury depending on anatomical conditions and clinical use. The deviation is limited to specific implant/sleeve/workflow selections and does not affect all protocol configurations.
What to do
Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.
Record last checked against the registry on 2026-05-31.