ProductGuru Recalls

Official recall record

Relay®Pro Thoracic Stent-Graft System

Health Canada — Medical Devices · 2026-05-29 · CanadaType I
Issuing authority
Health Canada — Medical Devices
Published
2026-05-29
Country of alert
Canada
Alert level
Type I
Alert type
recall

What the authority reported

Through post-market surveillance, Bolton Medical received complaints where the graft is unable to unclasp from the delivery system because the proximal clasp is disconnected from the outer control tube. This may be recognized by the user as a lack of resistance felt when sliding the apex holder back accompanied by a failure to release the proximal stent. The implant cannot be recaptured at this stage in the procedure. Bolton Medical received 4 complaints associated with this failure mode, 3 of which resulted in death including 1 aortic perforation, and 2 conversions to open surgery which resulted in patient deaths due to stroke.

What to do

Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.

Verify at the source: Health Canada — Medical Devices — official record →
Record last checked against the registry on 2026-06-10.

More recalls: Health Canada — Medical Devices · Canada · all filters

ProductGuru Recalls republishes official safety information with attribution. We are not the issuing authority and this page is not advice. If this product is a medical device or medication, speak to a doctor or pharmacist. Found an error? Report it — corrections are published.