IceMan CLASSIC3™

Health Canada — Medical Devices · 2026-06-03 · CanadaType II
Issuing authority
Health Canada — Medical Devices
Published
2026-06-03
Country of alert
Canada
Alert level
Type II
Alert type
recall

What the authority reported

Complaints received reporting a consistent failure mode described as a broken connector on IceMan CLASSIC™ cube unit. If the connector is broken, the unit may not function as intended, which could result in a temporary interruption of prescribed cold therapy until a replacement unit is provided. Leakage could saturate the barrier (dressings) and contact the surgical site with non-sterile water, which could in potential cold-related tissue injury or surgical site infection.