Giraffe™ OmniBed™ Carestation™

Health Canada — Medical Devices · 2026-06-04 · CanadaType II
Issuing authority
Health Canada — Medical Devices
Published
2026-06-04
Country of alert
Canada
Alert level
Type II
Alert type
recall

What the authority reported

GE HealthCare has become aware of a potential issue involving the canopy soffit beneath the control panel end of the device. Under certain conditions, for example, impact during transport, one or more of the fasteners that secure the soffit can disengage. If this occurs, a fastener could fall into the patient compartment. In rare circumstances, if multiple fasteners disengage, the canopy soffit can become loose or detach. These conditions could potentially result in patient injury.

What to do

Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.

Verify at the source: Health Canada — Medical Devices — official record →
Record last checked against the registry on 2026-06-14.