Harmony™ Delivery Catheter System

Health Canada — Medical Devices · 2026-06-04 · CanadaType II
Issuing authority
Health Canada — Medical Devices
Published
2026-06-04
Country of alert
Canada
Alert level
Type II
Alert type
recall

What the authority reported

Medtronic had received two (2) complaints about Harmony™ DCS distal tip detachment during the implant procedure. A root cause investigation has identified that, due to a manufacturing issue, Harmony™ DCS from specific manufacturing lots may have an increased potential for distal tip detachment under certain procedural or anatomical conditions.

What to do

Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.