Official recall record
RAPIDLab® 348EX System
- Issuing authority
- Health Canada
- Alert date
- 2026-07-02
- Country of alert
- Canada
- Alert level
- Type II
- Alert type
- recall
What the authority reported
Siemens Healthcare is informing affected customers that a software-related behavior identified in the RAPIDLab® 348EX System associated with the calcium ion concentration adjusted to H 7.4 result when the instrument is connected to a Laboratory Information System (LIS).
What to do
Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.
Verify at the source:
Health Canada — official record →
Record last checked against the registry on 2026-07-12.
Record last checked against the registry on 2026-07-12.
ProductGuru Recalls republishes official safety information with attribution. We are not
the issuing authority and this page is not advice. If this product is a medical device or
medication, speak to a doctor or pharmacist. Found an error?
Report it — corrections are published.