Official recall record
LIFECODES LSA Class II
- Issuing authority
- Health Canada
- Alert date
- 2026-07-02
- Country of alert
- Canada
- Alert level
- Type II
- Alert type
- recall
What the authority reported
Werfen has identified that two beads within the LIFECODES LSA class II assay kit(s) may present different DQB1 specificities than previously understood: • Bead 42 / Antigen ID 242 Previously associated with DQB1*02:01, this bead instead presents DQB1*02:02 • Bead 44 / Antigen ID 244 Previously associated with DQB1*02:02, this bead instead presents DQB1*02:01 • Note: DQA1*02:01 is present on both beads and therefore not impacted. Due to this issue, certain reactivity patterns may appear differently than expected.
What to do
Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.
Record last checked against the registry on 2026-07-12.