ProductGuru Recalls

Official recall record

Astral 150™

Health Canada · 2026-07-09 · CanadaType II
Issuing authority
Health Canada
Alert date
2026-07-09
Country of alert
Canada
Alert level
Type II
Alert type
recall

What the authority reported

An internal electrical component (supercapacitor) may leak electrolyte over time. In rare cases, this may damage specific circuitry on the ventilator's Printed Circuit Board Assembly ("PCBA"), causing the ventilator to inadvertently enter a fail-safe state. In this state, the ventilator is no longer able to deliver therapy. Alternative means of ventilation must be provided.

What to do

Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.

Verify at the source: Health Canada — official record →
Record last checked against the registry on 2026-07-18.

More recalls: Health Canada · Canada · all filters

ProductGuru Recalls republishes official safety information with attribution. We are not the issuing authority and this page is not advice. If this product is a medical device or medication, speak to a doctor or pharmacist. Found an error? Report it — corrections are published.