Official recall record
Astral 150™
- Issuing authority
- Health Canada
- Alert date
- 2026-07-09
- Country of alert
- Canada
- Alert level
- Type II
- Alert type
- recall
What the authority reported
An internal electrical component (supercapacitor) may leak electrolyte over time. In rare cases, this may damage specific circuitry on the ventilator's Printed Circuit Board Assembly ("PCBA"), causing the ventilator to inadvertently enter a fail-safe state. In this state, the ventilator is no longer able to deliver therapy. Alternative means of ventilation must be provided.
What to do
Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.
Verify at the source:
Health Canada — official record →
Record last checked against the registry on 2026-07-18.