SEMA 3

Health Canada · 2026-07-16 · CanadaType II
Issuing authority
Health Canada
Alert date
2026-07-16
Country of alert
Canada
Alert level
Type II
Alert type
recall

What the authority reported

When an ECG is aquired with DEFIGARD Touch 7 or ARGUS PRO LifeCare 3 with the newest software, there might be a discrepancy between the aquiring device and SEMA. This results from inadequate processing by SEMA to an already filtered signal. SEMA should be updated to version 25.10.

What to do

Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.

Verify at the source: Health Canada — official record →
Record last checked against the registry on 2026-07-26.

More recalls: Health Canada · Canada · all filters