ProductGuru Recalls

Official recall record

V-Loc™ 90 and 180 Suture Products

Health Canada · 2026-07-16 · CanadaType II
Issuing authority
Health Canada
Alert date
2026-07-16
Country of alert
Canada
Alert level
Type II
Alert type
recall

What the authority reported

On 29-Jan-2026 during a quality inspection at a manufacturing site, a V-Loc 180 unit was found with a hole in the breather pouch on the Tyvek side. The breather pouch serves as the sterile barrier system for the suture. Investigation determined the briller plate on the sealing station may be causing the hole. No complaints have been identified for this issue, but lots within the issue bounding have been released to Distribution Centers (DCs) and sold to customers. These released lots may also contain units with holes in the Tyvek side of the breather pouch.

What to do

Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.

Verify at the source: Health Canada — official record →
Record last checked against the registry on 2026-07-26.

More recalls: Health Canada · Canada · all filters

ProductGuru Recalls republishes official safety information with attribution. We are not the issuing authority and this page is not advice. If this product is a medical device or medication, speak to a doctor or pharmacist. Found an error? Report it — corrections are published.