Official recall record
Shimadzu X-ray High Voltage Generators
- Issuing authority
- Health Canada
- Alert date
- 2026-07-21
- Country of alert
- Canada
- Alert level
- Type II
- Alert type
- recall
What the authority reported
Shimadzu Corporation has identified a potential issue in the X-ray High Voltage Generator. Contacts of the electromagnetic contactors inside the generator may remain welded, allowing the electricity to flow even if the system was tuned off. This can cause overheating of the circuit, damage to surrounding parts and potentially produce smoke.
What to do
Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.