ProductGuru Recalls

Official recall record

Shimadzu X-ray High Voltage Generators

Health Canada · 2026-07-21 · CanadaType II
Issuing authority
Health Canada
Alert date
2026-07-21
Country of alert
Canada
Alert level
Type II
Alert type
recall

What the authority reported

Shimadzu Corporation has identified a potential issue in the X-ray High Voltage Generator. Contacts of the electromagnetic contactors inside the generator may remain welded, allowing the electricity to flow even if the system was tuned off. This can cause overheating of the circuit, damage to surrounding parts and potentially produce smoke.

What to do

Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.

Verify at the source: Health Canada — official record →
Record last checked against the registry on 2026-07-30.

More recalls: Health Canada · Canada · all filters

ProductGuru Recalls republishes official safety information with attribution. We are not the issuing authority and this page is not advice. If this product is a medical device or medication, speak to a doctor or pharmacist. Found an error? Report it — corrections are published.