ProductGuru Recalls

Official recall record

Blanketrol® II 115V and Blanketrol® III 115V

Health Canada · 2026-08-28 · CanadaType II
Issuing authority
Health Canada
Alert date
2026-08-28
Country of alert
Canada
Alert level
Type II
Alert type
recall

What the authority reported

Gentherm Medical has received complaints of overheated circuit boards in 0.53% of distributed devices. Overheating can cause the circuit board to stop operating, which may result in a device that shuts down at startup or during use on a patient. Based on the low probability of harm, and absence of reported harm, associated with this issue, users may continue to use the device. In an abundance of caution, it is recommended to have an alternative means of providing therapy available.

What to do

Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.

Verify at the source: Health Canada — official record →
Record last checked against the registry on 2026-09-07.

More recalls: Health Canada · Canada · all filters

ProductGuru Recalls republishes official safety information with attribution. We are not the issuing authority and this page is not advice. If this product is a medical device or medication, speak to a doctor or pharmacist. Found an error? Report it — corrections are published.