ProductGuru Recalls

Official recall record

1788 4K Camera

Health Canada · 2026-08-05 · CanadaType II
Issuing authority
Health Canada
Alert date
2026-08-05
Country of alert
Canada
Alert level
Type II
Alert type
recall

What the authority reported

Stryker Endoscopy has received an increase in complaints for 1788 out of box failures due to bent connection pins. The bent pin may also damage the mating connector in the CCU. Stryker has investigated and identified a service repair issue which contributed to the bent pins noted with these devices.

What to do

Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.

Verify at the source: Health Canada — official record →
Record last checked against the registry on 2026-08-16.

More recalls: Health Canada · Canada · all filters