ProductGuru Recalls

Official recall record

StatStrip® Glucose And Beta-Ketone Hospital Meter

Health Canada · 2026-08-19 · CanadaType II
Issuing authority
Health Canada
Alert date
2026-08-19
Country of alert
Canada
Alert level
Type II
Alert type
recall

What the authority reported

A software issue in StatStrip (Gen 2) Glucose and Beta-Ketone Hospital Meters may retain previous patient information during barcode scanning, potentially assigning results to the wrong patient. Users should verify patient identification before accepting scanned data and follow the correction instructions provided by Nova Biomedical.

What to do

Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.

Verify at the source: Health Canada — official record →
Record last checked against the registry on 2026-08-29.

More recalls: Health Canada · Canada · all filters

ProductGuru Recalls republishes official safety information with attribution. We are not the issuing authority and this page is not advice. If this product is a medical device or medication, speak to a doctor or pharmacist. Found an error? Report it — corrections are published.