Official recall record
SPECTRALIS® with Flex Module
- Issuing authority
- Health Canada
- Alert date
- 2026-08-19
- Country of alert
- Canada
- Alert level
- Type II
- Alert type
- recall
What the authority reported
There is a possibility that the castor wheels of the Flex stand are not properly attached. The wheels may come loose. As a result, the Flex stand may tip over. Users shall schedule a service appointment as soon as possible so that the castor fastenings can be inspected and corrected.