Official recall record
iQ200 Series Analyzers and DxU 840m/850m Iris Analyzers
- Issuing authority
- Health Canada
- Alert date
- 2026-08-21
- Country of alert
- Canada
- Alert level
- Type II
- Alert type
- recall
What the authority reported
Beckman Coulter received customer complaints of lamina fluid leaks in iQ200 Series and DxU 840m/850m Analyzers. The investigation found that a structurally compromised rinse-line check valve may allow fluid to migrate internally and leak onto the table or floor, potentially resulting into exposure to physical or mechanical or chemical hazards.
What to do
Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.
Verify at the source:
Health Canada — official record →
Record last checked against the registry on 2026-09-02.
Record last checked against the registry on 2026-09-02.
ProductGuru Recalls republishes official safety information with attribution. We are not
the issuing authority and this page is not advice. If this product is a medical device or
medication, speak to a doctor or pharmacist. Found an error?
Report it — corrections are published.