Dxc 500i Clinical Analyzer and Dxc 500 AU Clinical Chemistry Analyzer
- Issuing authority
- Health Canada
- Alert date
- 2026-09-10
- Country of alert
- Canada
- Alert level
- Type II
- Alert type
- recall
What the authority reported
Beckman Coulter identified a software issue affecting DxC 500i and DxC 500 AU Analyzers when testing plasma samples. Following user configuration of a chemistry test to nondefault units, subsequent activation of a draft configuration in the system configuration page may revert units to the system default without notifying the user.