Official recall record
IntelliSpace Critical Care And Anesthesia
- Issuing authority
- Health Canada
- Alert date
- 2026-09-16
- Country of alert
- Canada
- Alert level
- Type II
- Alert type
- recall
What the authority reported
In ICCA releases J.06.01 and J.06.02, calculations triggered by infusion-pump auto-charting may result in incorrect Intake Totals, Output Totals, Net Body Balance, Length of Stay (LOS) Balance, and related fluid-balance values. This issue applies only to customers that have configured and use Auto Charting for Infusion Data for a Clinical Unit, with one or both of the following settings enabled: • Automatic periodic charting of infusion rates and volumes. • Automatic aperiodic charting of infusion rates and bolus volumes when a change occurs. Customers that do not have either Auto Charting for Infusion Data setting configured and in use are not impacted by this issue.