SIGNA™ Artist and SIGNA™ Artist Evo
- Issuing authority
- Health Canada
- Alert date
- 2026-09-18
- Country of alert
- Canada
- Alert level
- Type II
- Alert type
- recall
What the authority reported
GE Healthcare has become aware of an exposed sharp edge inside the SIGNA™ Artist Evo and SIGNA™ Artist electronics cabinet that could damage power cables, which may potentially cause an electrical fault. If the electrical fault is present and the cabinet door is opened while the system is powered, this could potentially result in an arc flash that could lead to serious injury to service personnel in immediate proximity.
What to do
Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.