ProductGuru Recalls

Official recall record

PERMA-HAND™ Silk Suture

Health Canada · 2026-09-25 · CanadaType II
Issuing authority
Health Canada
Alert date
2026-09-25
Country of alert
Canada
Alert level
Type II
Alert type
recall

What the authority reported

Ethicon determined that PERMA-HAND™ Silk Suture subject to this recall (removal) was incorrectly manufactured with an alcohol-based tubing fluid in the primary packaging. This tubing fluid is an isopropyl alcohol (IPA), diethylaminoethanol (DEAE), and sodium benzoate solution previously used in the manufacture of a different suture product. Up to 80% of the impacted lot may have had the alcohol-based tubing fluid. Tubing fluid is not a material approved for use in the manufacture of PERMA-HAND™ Silk Suture.

What to do

Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.

Verify at the source: Health Canada — official record →
Record last checked against the registry on 2026-09-29.

More recalls: Health Canada · Canada · all filters