EMAG® Instrument

Health Canada — Medical Devices · 2026-10-02 · CanadaType II
Issuing authority
Health Canada — Medical Devices
Published
2026-10-02
Country of alert
Canada
Alert level
Type II
Alert type
recall

What the authority reported

bioMérieux received complaints from the field related to EMAG® system Reference 418591 Error 10003 Heater timeout issues. The risk associated with the referenced issue is a potential delayed result / no result (loss of sample). Initial investigation of the issue identified the error started to occur on the field after firmware SP4 Patch 2 was installed. The investigation is ongoing to identify the root cause of the 10003-error heater timeout issue.