Official recall record
Radiography 7000 M
- Issuing authority
- Health Canada — Medical Devices
- Published
- 2026-10-01
- Country of alert
- Canada
- Alert level
- Type II
- Alert type
- recall
What the authority reported
Philips has become aware of a potential safety issue impacting the Radiography 7000 M system (Model 712010) where the drive function may fail immediately after an exposure, causing the system to remain stationary. The issue may occur with the Radiography 7000 M system immediately after an exposure, causing the drive function to fail and preventing repositioning of the system. The system may remain stationary instead of driving upon activation of the handlebar. The issue is intermittent and has been observed during routine mobile radiography use.
What to do
Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.
Record last checked against the registry on 2026-10-12.