CareLink™ SmartSync™ Device Manager (ref 82715)
- Issuing authority
- Health Canada — Medical Devices
- Published
- 2026-10-01
- Country of alert
- Canada
- Alert level
- Type II
- Alert type
- recall
What the authority reported
A complaint from the CareLink network reported a sudden decrease in the Remaining Longevity of a patient's Micra VR CIED from 3 years to 6 months following a remote transmission, which was later found to be incorrect. Analysis revealed a calculation issue in SmartSync and CareLink for Micra VR2 and Micra AV2 devices, occurring under specific conditions. This miscalculation stems from a software error where a function received an integer instead of a real data type, defaulting the average pacing rate to 60 BPM and resulting in inaccurate Remaining Longevity estimates.