Official recall record
Lumify Ultrasound System
- Issuing authority
- Health Canada — Medical Devices
- Published
- 2026-10-05
- Country of alert
- Canada
- Alert level
- Type I
- Alert type
- recall
What the authority reported
Potential issues with the Lumify Diagnostic Ultrasound System which may prevent or delay scanning functionality: Due to a software issue, the Lumify application may not recognize a previously registered transducer and prompt for re-registration (outside of the defined criteria in the user guide) without warning prior to point of care use. Scanning functionality is prevented until the registration step has been completed; registration requires wi-fi or cellular signal access. Lumify devices used outside the hospital, such as in ambulances, aircraft, or field settings, are at greater risk due to reduced access to wi-fi/cellular. Philips has received reports that these erroneous re-registration events have prevented scanning in pre-hospital environments and delayed care to patients. These pre-hospital events occurred outside of wi-fi or cellular access. Additionally, Philips has identified that the connection between the Lumify transducer USB cable to a customer's tablet/ phone may disconnect or have an unstable connection during use which may disrupt continuous function of the device and result in a delay of care.
What to do
Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.
Record last checked against the registry on 2026-10-15.