ProductGuru Recalls

Official recall record

Avantage Inserter Handles and Grippers

Health Canada — Medical Devices · 2026-10-08 · CanadaType II
Issuing authority
Health Canada — Medical Devices
Published
2026-10-08
Country of alert
Canada
Alert level
Type II
Alert type
recall

What the authority reported

It was identified while using the Avantage inserter handles and grippers during the implantation of the acetabular shell, that the acetabular shell may not consistently disengage from the gripper when the trigger is unlocked. As a result, the inserter handle and gripper assembly may remain engaged with the implanted shell, preventing clean separation of the instrumentation from the implant after its seating.

What to do

Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.

Verify at the source: Health Canada — Medical Devices — official record →
Record last checked against the registry on 2026-10-17.

More recalls: Health Canada — Medical Devices · Canada · all filters