cobas® c 303 analytical unit

Health Canada — Medical Devices · 2026-10-15 · CanadaType II
Issuing authority
Health Canada — Medical Devices
Published
2026-10-15
Country of alert
Canada
Alert level
Type II
Alert type
recall

What the authority reported

An issue has been identified where vacuum tubing of the cell rinse unit of the cobas c 303 analytical unit prematurely deteriorated. Investigations confirmed the issue. In the event of small pinholes or cracks, no impact on analytical performance was identified. However, in case of tubing rupture, the following can occur as a result of the issue: Undetected leakage: Fluid may drip into random reaction cells. Result impact: Increased risk of erroneous patient and QC results.