i-STAT hs-TnI Cartridge

Health Canada — Medical Devices · 2026-10-21 · CanadaType II
Issuing authority
Health Canada — Medical Devices
Published
2026-10-21
Country of alert
Canada
Alert level
Type II
Alert type
recall

What the authority reported

Abbott has identified that lots beginning with A24, A25, B25, B26, C26, and D26 may have a reduced ability to mitigate certain forms of analytical interference, including macrotroponin and heterophile antibody-related interference, due to variability associated with the alkaline phosphatase component used during reagent manufacturing. This issue may result in falsely elevated cardiac troponin I results. There is no evidence that the issue has caused falsely low results.