i-STAT hs-TnI Cartridge
- Issuing authority
- Health Canada — Medical Devices
- Published
- 2026-10-21
- Country of alert
- Canada
- Alert level
- Type II
- Alert type
- recall
What the authority reported
Abbott has identified that lots beginning with A24, A25, B25, B26, C26, and D26 may have a reduced ability to mitigate certain forms of analytical interference, including macrotroponin and heterophile antibody-related interference, due to variability associated with the alkaline phosphatase component used during reagent manufacturing. This issue may result in falsely elevated cardiac troponin I results. There is no evidence that the issue has caused falsely low results.
What to do
Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.