PD-L1 IHC 22C3 pharmDx

Health Canada — Medical Devices · 2026-10-28 · CanadaType II
Issuing authority
Health Canada — Medical Devices
Published
2026-10-28
Country of alert
Canada
Alert level
Type II
Alert type
recall

What the authority reported

Agilent identified missing hyphens in the IFU for PD-L1 IHC 22C3 pharmDx (GE006), revision 02 (PD005650). Due to an inadvertent formatting change in section "9.2 tissue sections," tissue-section thickness was incorrectly listed as 45 μm instead of 4-5 μm, and slide-storage temperature was incorrectly listed as 28 °c instead of 2-8 °c. This issue is limited to the electronic IFU (eIFU) associated with the three lots.