ProductGuru Recalls

Official recall record

Abre Venous Self-Expanding Stent System

Health Canada — Medical Devices · 2026-10-26 · CanadaType II
Issuing authority
Health Canada — Medical Devices
Published
2026-10-26
Country of alert
Canada
Alert level
Type II
Alert type
recall

What the authority reported

Medtronic is initiating this recall to inform customers of an update to the Instructions For Use (IFU) for the Abre™ Venous Self-Expanding Stent System (herein referred to as Abre). This update provides a validated manual deployment workaround method to help mitigate potential harms in the rare event customers experience a partial stent deployment; the method is included in the notification letter and the Abre IFU will be updated, pending regulatory approvals, to reflect this standardized approach. These instructions should only be followed in the event a partial stent deployment occurs. Medtronic is taking this action to reduce the severity of potential patient harms. The steps provided in the notification are intended to minimize the risk of serious adverse health consequences in the rare event of partial stent deployment. There are no product retrievals or disposals requested by Medtronic as part of this IFU update.

What to do

Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.

Verify at the source: Health Canada — Medical Devices — official record →
Record last checked against the registry on 2026-11-05.

More recalls: Health Canada — Medical Devices · Canada · all filters