ProductGuru Recalls

Official recall record

BIOFIRE Blood Culture Identification 2 (BCID2), 30-Pack

Health Canada — Medical Devices · 2026-10-28 · CanadaType II
Issuing authority
Health Canada — Medical Devices
Published
2026-10-28
Country of alert
Canada
Alert level
Type II
Alert type
recall

What the authority reported

BioMérieux has received complaints of false negative results on BIOFIRE BCID2 Panel kit lot 0727226 / pouch lot 3ZQT26 and investigation identified a product issue. This field safety notice is issued due to unintended movement of pouch chemistry in BIOFIRE BCID2 Panel kit lot 0727226, which based on investigation, can lead to an elevated risk of false negative results. All analytes on the panel may be at risk of false negative results. No other BIOFIRE product lots were determined to be impacted.

What to do

Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.

Verify at the source: Health Canada — Medical Devices — official record →
Record last checked against the registry on 2026-11-07.

More recalls: Health Canada — Medical Devices · Canada · all filters

ProductGuru Recalls republishes official safety information with attribution. We are not the issuing authority and this page is not advice. If this product is a medical device or medication, speak to a doctor or pharmacist. Found an error? Report it — corrections are published.