ProductGuru Recalls

Official recall record

RotaFlow Console

Health Canada — Medical Devices · 2026-10-29 · CanadaType II
Issuing authority
Health Canada — Medical Devices
Published
2026-10-29
Country of alert
Canada
Alert level
Type II
Alert type
recall

What the authority reported

During internal validation of new replacement batteries, MCP identified an issue on the RotaFlow console. This issue can cause the device to stop battery charging after the battery operation mode has been utilized. Further investigation identified a software problem as root cause. Because of this, all devices on the market are considered affected by this issue.

What to do

Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.

Verify at the source: Health Canada — Medical Devices — official record →
Record last checked against the registry on 2026-11-08.

More recalls: Health Canada — Medical Devices · Canada · all filters