Official recall record
Horizon Medical Imaging System
- Issuing authority
- Health Canada — Medical Devices
- Published
- 2026-11-05
- Country of alert
- Canada
- Alert level
- Type II
- Alert type
- recall
What the authority reported
It was identified that certain replacement monitors for Horizon W/WI systems released into production beginning in January 2025 were shipped without the VGA cable required to connect the monitor to the computer. As a result, field service engineers may have installed the systems using VGA cables from unknown origin. This created a nonconformance related to the use of unapproved components and a potential risk of increased electromagnetic interference (EMI) emissions from VGA cables that have not been evaluated or approved by Hologic.