Official recall record
Smoothbore Breathing System
- Issuing authority
- Health Canada — Medical Devices
- Published
- 2026-11-06
- Country of alert
- Canada
- Alert level
- Type II
- Alert type
- recall
What the authority reported
During manufacturing, insufficient tetrahydrofuran (THF) solvent is used to bond the soft connectors onto the Smoothbore tube. Following reports received from the market of disconnection of the soft connectors from the tube, our investigation and testing of internal stock has identified insufficient amount of solvent used to ensure a secure connection.
What to do
Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.
Verify at the source:
Health Canada — Medical Devices — official record →
Record last checked against the registry on 2026-11-13.
Record last checked against the registry on 2026-11-13.
ProductGuru Recalls republishes official safety information with attribution. We are not
the issuing authority and this page is not advice. If this product is a medical device or
medication, speak to a doctor or pharmacist. Found an error?
Report it — corrections are published.