Official recall record
700hpm high pressure calibration manifold
- Issuing authority
- OPSS (UK)
- Alert date
- 2024-07-31
- Country of alert
- United Kingdom
- Alert level
- Product Recall
- Alert type
- product-recall
- Brand on record
- Fluke
What the authority reported
This product presents a serious risk of injuries as during manual pressurization of the equipment, the plunger nut threads of the High-Pressure Pneumatic Calibration Manifold may yield releasing the plunger assembly and allowing it to be ejected at high velocity. The threads securing the plunger assembly have been found to fail prematurely. This can result in serious injury to the user. The product does not meet the requirements of the Pressure Equipment (Safety) Regulations 2016.