Official recall record
Beha amprobe mt204 s with test leads
- Issuing authority
- OPSS (UK)
- Alert date
- 2024-04-15
- Country of alert
- United Kingdom
- Alert level
- Product Recall
- Alert type
- Product recall
What the authority reported
EN 61557-14, requires two-hand operation for test equipment when used for high voltage testing as an additional level of protection against electric shock during use. As of 1 July 2018, the shrouded test leads set initially supplied for use with the Machinery Tester, which required two-hand use, was inadvertently replaced with an alternative unshrouded test leads set that did not require two-hand operation. The inclusion of unshrouded test leads resulted in the machinery tester not being compliant with EN 61577-14 and not compliant with the essential health and safety requirements.
What to do
Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.