Product Wiki · Recalls · U.S. FDA — Device Enforcement

Zenition 50, Model #718096 — recall Z-0647-2022

U.S. FDA — Device Enforcement · Netherlands · 2021-12-02 · Class II

Compiled from an official source — the facts on this page are reproduced from the issuing authority’s own published notice, which is linked on this page. No AI-generated text appears on this page. How we use AI →
Agency
U.S. FDA — Device Enforcement
Recall identifier
Z-0647-2022
Date of alert
2021-12-02
Country / jurisdiction
Netherlands
Risk level
Class II
Notice type
recall
Brand
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
Product category
device
Product type
Devices
Recall status
Ongoing — the recall is still in progress
Initiated by
Voluntary: Firm initiated
Distribution
Domestic distribution to AZ CA FL GA HI IL KS KY MA MD ME MI MN MS NC NE NJ NV NY OH PA TN TX VA WA WV. Worldwide foreign distribution.
Reported risk: Firm discovered that the wireless foot switch can suddenly stop responding due to a firmware issue. If the wireless foot switch fails, there could be a delay or interruption of procedure if the hand switch is not used or if the wired foot switch is not connected to the system.
Primary source: https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfRes/res.cfm?start_search=1&

Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.

ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.