Product Wiki · Recalls · U.S. FDA — Device Enforcement

Philips Allura Xper Systems (R7.6-R8.1) — recall Z-0676-2025

U.S. FDA — Device Enforcement · Netherlands · 2024-11-11 · Class II

As described by U.S. FDA — Device Enforcement: Philips Allura Xper Systems (R7.6-R8.1). Labeled as the following: 1. Allura Xper FD10, Model Number: 722010. 2. Allura Xper FD10/10, Model Number: 722011. 3. Allura Xper FD20, Model Number: 722012. 4. Allura Xper FD20 Biplane, Model Number: 722013. 5. Allura Xper FD10 OR Table, Mode

Compiled from an official source — the facts on this page are reproduced from the issuing authority’s own published notice, which is linked on this page. No AI-generated text appears on this page. How we use AI →
Agency
U.S. FDA — Device Enforcement
Recall identifier
Z-0676-2025
Date of alert
2024-11-11
Country / jurisdiction
Netherlands
Risk level
Class II
Notice type
recall
Brand
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
Product category
device
Product type
Devices
Recall status
Ongoing — the recall is still in progress
Initiated by
Voluntary: Firm initiated
Distribution
Nationwide distribution. International distribution to Albania, Algeria, Argentina, Australia, Austria, Azerbaijan, Bangladesh, Belgium, Brazil, Bulgaria, Canada, Cayman Islands, Chile, China, Colombia, Costa Rica, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, Estonia, Finland, Franc
Barcode (EAN/GTIN)
00884838054189
Reported risk: Potential safety issues in Allura Xper R7.6- R8.1 systems where the Lifetime Extension (LTE) kit is not installed in a timely manner. Root cause is under investigation, but the recall reason is listed as for systems over 10 years of life with LTE kits are experiencing propeller motor clamping bolt and x-ray tube locking bolt breaks. If this happens, it may cause erratic C-arm movements and potential collision with patients or bystanders, or detection of a false collisions and subsequent blocked movement of the c-arm. Movement may also cause image noise and black area on the x-ray image, and possible need to expose the patient to additional unnecessary radiation during re-examination.
What to do

Check whether your product matches this notice (name, batch, dates). If it does, stop using it and follow the instructions in the official notice — refund, return, repair or disposal are set by the issuing authority and the business, and they differ per recall. Official notice from U.S. FDA — Device Enforcement →

Primary source: https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfRes/res.cfm?start_search=1&

Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.

ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.