ProductGuru Recalls

Official recall record

Multichem S Plus, Part Codes: a) 08P88-10, b) 08P88-11, c) 08P88-12

US FDA — Medical Devices · 2023-08-25 · IrelandClass II
Issuing authority
US FDA — Medical Devices
Recall initiated by firm
2023-08-25
Country of alert
Ireland
Alert level
Class II
Alert type
recall
Barcode (EAN/GTIN)
05391523440776
Brand on record
TECHNO-PATH MANUFACTURING LTD.

What the authority reported

Technopath Manufacturing Ltd. has identified an issue with the glass vial commodity used to fill the Multichem S Plus Level 1, Multichem S Plus Level 2, Multichem S Plus Level 3, Multichem IA Plus and Multichem P Controls. Some customers have experienced broken vials when receiving or thawing Technopath Multichem IA Plus and Multichem S Plus Controls. Breakages have been observed upon delivery, while thawing or while opening the vial. The breakages can result in the bottom of the vial breaking off, breakage in the neck thread area or cracking on the sides of the vial. This may expose the user to biohazardous material if cut and may cause a delay in patient results.

What to do

Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.

Verify at the source: US FDA — Medical Devices — official record →
Record last checked against the registry on 2026-05-31.

More recalls: US FDA — Medical Devices · Ireland · all filters

Other recalls from this brand

All 5 official recall notices naming TECHNO-PATH MANUFACTURING LTD. on this site →

ProductGuru Recalls republishes official safety information with attribution. We are not the issuing authority and this page is not advice. If this product is a medical device or medication, speak to a doctor or pharmacist. Found an error? Report it — corrections are published.