Nalu TD2 Therapy Disc with firmware, REF: 34014-001, and TD2 Patient Kit, REF: 74011, part of the Nalu Neurostimulation System

US FDA — Medical Devices · 2026-09-30 · United StatesClass II
Issuing authority
US FDA — Medical Devices
Recall initiated by firm
2026-09-30
Country of alert
United States
Alert level
Class II
Alert type
recall
Barcode (EAN/GTIN)
00812537037134
Brand on record
Boston Scientific Neuromodulation Corporation

What the authority reported

Therapy Disc (TD) that powers and sends stimulation signals to an Implantable Pulse Generator (IPG), part of a neurostimulation system, has a firmware issue that causes the TD to attempt to restart, then fail, leading it to become non-functional, and if a secondary TD is unavailable, this may lead IPG to become non-functional, thus unable to inhibit transmission of pain signals.

What to do

Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.

Verify at the source: US FDA — Medical Devices — official record →