ELEKTA LIMITED Elekta Infinity, REF XRT 0511, XRT 0521, XRT 0531, XRT 0541; Medical Charged-Particle Radiation Therapy System Accelerator, L

US FDA — Medical Devices · 2023-10-25 · United StatesClass II
Issuing authority
US FDA — Medical Devices
Recall initiated by firm
2023-10-25
Country of alert
United States
Alert level
Class II
Alert type
recall
Barcode (EAN/GTIN)
05060191071512
Brand on record
Elekta, Inc.

What the authority reported

Elekta has identified that if the microswitch within the middle arm is incorrectly setup, there is a potential for an uncontrolled extension of iViewGT / XVI detector arms. When the detector arm is fully deployed there is no risk of uncontrolled extension.

What to do

Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.

Verify at the source: US FDA — Medical Devices — official record →
Record last checked against the registry on 2026-05-31.