BIOFIRE Blood Culture Identification 2 (BCID2) Panel, REF: RFIT-ASY-0147, used with Maine Molecular Quality Controls, Inc (MMQCI) FilmArray
- Issuing authority
- US FDA — Medical Devices
- Recall initiated by firm
- 2023-10-17
- Country of alert
- United States
- Alert level
- Class II
- Alert type
- recall
- Barcode (EAN/GTIN)
- 00815381020338
- Brand on record
- BioFire Diagnostics, LLC
What the authority reported
Routine quality control testing of affected blood culture identification panel with affected control panel may need to be amended, due to control panel/instrument characteristic changes, that lead C. tropicalis to be amplified in some portion of test runs, but then it's reported as "Not Detected" because the assay amplicon Tm is outside the acceptable limit assigned for calling positive results.
What to do
Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.
Record last checked against the registry on 2026-05-31.
Same barcode in other registries
The same EAN appears in 2 other official recall record(s) — a cross-border signal worth checking if you bought this product abroad.
- 2024-06-19 · US FDA — Medical Devices
BIOFIRE Blood Culture Identification 2 (BCID2) Panel, REF: RFIT-ASY-0147, for FILMARRAY Systems, when used in conjunctio - 2021-10-11 · US FDA — Medical Devices
BioFire Blood Culture Identification 2 (BCID2) Panel Part No: RFIT-ASY-0147 UDI: 00815381020338 Blood Culture Catalog N
Other recalls from this brand
- 2024-08-07 · US FDA — Medical Devices
FILMARRAY 2.0 INSTRUMENT, REF: FLM2-ASY-0001, FLM2-ASY-0001-W