Official recall record
FILMARRAY 2.0 INSTRUMENT, REF: FLM2-ASY-0001, FLM2-ASY-0001-W
- Issuing authority
- US FDA — Medical Devices
- Alert date
- 2024-08-07
- Country of alert
- United States
- Alert level
- Class II
- Alert type
- recall
- Barcode (EAN/GTIN)
- 00815381020031
- Brand on record
- BioFire Diagnostics, LLC
What the authority reported
Some automated in vitro diagnostic modules and instruments manufactured or serviced between May 2022 and May 2024 were calibrated with internal equipment that may lead modules and instruments to report higher melting temperatures, which may lead to erroneous results including false positive Salmonella, Giardia lamblia, false positive/negative Norovirus, and false negative MRSA.
What to do
Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.
Verify at the source:
US FDA — Medical Devices — official record →
Record last checked against the registry on 2026-05-31.
Record last checked against the registry on 2026-05-31.