ProductGuru Recalls

Official recall record

RAPIDPoint 500 Systems Measurement Cartridge (with Lactate) 400-Intended for in vitro diagnostic use and is designed to provide the determin

US FDA — Medical Devices · 2023-10-26 · United StatesClass II
Issuing authority
US FDA — Medical Devices
Recall initiated by firm
2023-10-26
Country of alert
United States
Alert level
Class II
Alert type
recall
Barcode (EAN/GTIN)
00630414589763
Brand on record
Siemens Healthcare Diagnostics Inc

What the authority reported

RAPIDPoint 500 Systems Measurement Cartridges (with lactate) has the potential to affect the Sodium (Na+) sensor to produce a negative bias that could result in serious injury due to a delayed diagnosis of hypernatremia or unnecessary intervention for hyponatremia and result in iatrogenic hypernatremia especially if the true sodium result is near the thresholds of severe alterations, as well as cause a Question Result -----? error flag for multiple electrolytes on patient samples and quality control that may cause a delay of diagnosis or treatment.

What to do

Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.

Verify at the source: US FDA — Medical Devices — official record →
Record last checked against the registry on 2026-05-31.

More recalls: US FDA — Medical Devices · United States · all filters

Other recalls from this brand

All 38 official recall notices naming Siemens Healthcare Diagnostics, Inc. on this site →