Scoot Control R-net, Product Number P015-61. The Scoot Control is intended to support a wheelchair attendant in controlling or maneuvering a

US FDA — Medical Devices · 2023-11-28 · BelgiumClass II
Issuing authority
US FDA — Medical Devices
Recall initiated by firm
2023-11-28
Country of alert
Belgium
Alert level
Class II
Alert type
recall
Barcode (EAN/GTIN)
05407008320119
Brand on record
mo-Vis BVBA

What the authority reported

In rare cases, the lever of the throttle can get temporarily stuck during the movement and does not come back to its neutral point by itself. If this happens during usage, an unwanted movement of the chair could be the result. An unwanted movement could result in damage to the wheelchair, material damage around the chair and/or injury to the user, attendant or by-standers.

What to do

Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.