ProductGuru Recalls

Official recall record

AirLife Adult Manual Resuscitator, Variable Volume Oxygen Reservoir Tubing, Adult Mask, REF 2K8017

US FDA — Medical Devices · 2024-01-10 · United StatesClass I
Issuing authority
US FDA — Medical Devices
Recall initiated by firm
2024-01-10
Country of alert
United States
Alert level
Class I
Alert type
recall
Barcode (EAN/GTIN)
50190752114294
Brand on record
Vyaire Medical
Product category
Medical Device

What the authority reported

Risk of hypoventilation caused by duckbill and ring broken/disassembled, due to faulty mold/tooling producing out of specification components. Device failure may cause improper ventilation or no ventilation that may result in hypoventilation or hypoxia, and potentially lead to death. The mold/tooling used to produce the affected component was replaced in 2017.

What to do

Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.

Verify at the source: US FDA — Medical Devices — official record →
Record last checked against the registry on 2026-05-31.

More recalls: US FDA — Medical Devices · United States · all filters

Other recalls from this brand

All 6 official recall notices naming Vyaire Medical on this site →

ProductGuru Recalls republishes official safety information with attribution. We are not the issuing authority and this page is not advice. If this product is a medical device or medication, speak to a doctor or pharmacist. Found an error? Report it — corrections are published.