Official recall record
Paradigm insulin pump, REF: MMT-754
- Issuing authority
- US FDA — Medical Devices
- Recall initiated by firm
- 2025-01-31
- Country of alert
- United States
- Alert level
- Class II
- Alert type
- recall
- Brand on record
- Medtronic MiniMed, Inc.
What the authority reported
All Medtronic MiniMed pumps (Paradigm series, Minimed 530G, 6XX series, and 7XX series) were found to have the potential for abnormal insulin delivery due to changes in air pressure, which can be found during airplane take-off or landing. An unexpected additional dose of insulin, which can occur during take-off, can result in severe hypoglycemia, altered mental status, seizure, coma, or death. An unexpected under-delivery of insulin, which can occur during landing, can result in severe hyperglycemia, dehydration, diabetic ketoacidosis, or death.
What to do
Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.
Record last checked against the registry on 2026-05-31.
Other recalls from this brand
- 2026-08-20 · US FDA — Medical Devices
Brand Name: MiniMed Flex Product Name: MiniMed Flex Insulin Pump Model/Catalog Numbers: MMT-8062; MMT-8072 Pro - 2026-07-13 · US FDA — Medical Devices
MiniMed 640G Insulin Pump, Model: MMT-1711, MMT-1712; - 2026-07-13 · US FDA — Medical Devices
MiniMed 670G Insulin Pump, Model: MMT-1781, MMT-1782;
All 30 official recall notices naming Medtronic MiniMed, Inc. on this site →