MiniMed 530G Insulin Pump (MMT-551, MMT-751)
- Issuing authority
- US FDA — Medical Devices
- Recall initiated by firm
- 2026-02-13
- Country of alert
- United States
- Alert level
- Class II
- Alert type
- recall
- Barcode (EAN/GTIN)
- 00643169507784
- Brand on record
- Medtronic MiniMed, Inc.
What the authority reported
All Medtronic MiniMed infusion pumps (Paradigm series, 600 series, and BLE 700 series) were found to be affected by unintended over- and under-delivery of insulin when the pump is elevated or lowered relative to the infusion site, respectively, because of changes in gravitational force impacting hydrostatic and hydrodynamic pressures. Insulin over-delivery, which can occur when the pump is elevated relative to the infusion site, can result in severe hypoglycemia, altered mental status, seizure, coma, or death. An under-delivery of insulin, which can occur at lowered pump height conditions relative to the infusion site, can result in severe hyperglycemia, dehydration, diabetic ketoacidosis, or death.
What to do
Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.
Record last checked against the registry on 2026-05-31.
Other recalls from this brand
- 2026-08-20 · US FDA — Medical Devices
Brand Name: MiniMed Flex Product Name: MiniMed Flex Insulin Pump Model/Catalog Numbers: MMT-8062; MMT-8072 Pro - 2026-07-13 · US FDA — Medical Devices
MiniMed 640G Insulin Pump, Model: MMT-1711, MMT-1712; - 2026-07-13 · US FDA — Medical Devices
MiniMed 670G Insulin Pump, Model: MMT-1781, MMT-1782;
All 30 official recall notices naming Medtronic MiniMed, Inc. on this site →